Posted by Boomer on October 2, 2005, 10:16 am
Please log in for more thread options : 7bit
Recently we have been worried about reefers using and assuming that the refracts
that ever
one has been using in this hobby should be off - 1.5 ppt or -.15 % or - .0015
SG. We have
been telling people in short to raise the seamix "15 units" higher to = 35 ppt
( to read
36.5), 3.5 ( to read 3.65 and 1.0265 A( to read 1,0280).
These tests have been completed on a number BUT the SAME brand refracts. These
are the
ones on the below image. Low and behold, they should read these values on a good
lab
refract and would need these adjustments but about 95 % of the ones reefers are
using are
off from the factory that exact amount we are correcting for :lol:
So it appears that with the these Chinese refracts, zeroed to RO/DI water, that
they are
right on when tested to a seawater std, when they should not be. So it reads 35
ppt, 3.5 %
or 1,0265 no adjustment is needed to our surprise. Just use as is :D
Note: This is only for these refracts !!
RHS-10ATC
http://www.instrument-china.com/images/32-t.gif
--
Boomer
Want to talk chemistry ? The Reef Chemistry Forum
http://www.reefcentral.com/vbulletin/index.php
WCWing@nospamChartermi.Net
Former US Army Bomb Technician (EOD)
Member; IABTI, NATEODA, WEODF, ISEE & IPS
If You See Me Running You Better Catch-Up
------=_NextPart_000_0012_01C5C731.F0F2DC20
Content-Type: text/html;
charset="iso-8859-1"
Content-Transfer-Encoding: quoted-printable
<!DOCTYPE HTML PUBLIC "-//W3C//DTD HTML 4.0 Transitional//EN">
<HTML><HEAD>
<META http-equiv=3DContent-Type content=3D"text/html; =
charset=3Diso-8859-1">
<META content=3D"MSHTML 6.00.2900.2096" name=3DGENERATOR>
<STYLE></STYLE>
</HEAD>
<BODY bgColor=3D#ffffff>
<DIV><FONT face=3DArial size=3D2>
<DIV><FONT face=3DArial size=3D2>Recently we have been worried about =
reefers using=20
and assuming that the refracts that ever one has been using in this =
hobby should=20
be off - 1.5 ppt or -.15 % or - .0015 SG. We have been telling people in =
short=20
to raise the seamix "15 units" higher to =3D 35 ppt ( to read =
36.5), 3.5 (=20
to read 3.65 and 1.0265 A( to read 1,0280). </FONT></DIV>
<DIV> </DIV>
<DIV><FONT face=3DArial size=3D2>These tests have been completed on a =
number BUT the=20
SAME brand refracts. These are the ones on the below image. Low and =
behold, they=20
should read these values on a good lab refract and would need these =
adjustments=20
but about 95 % of the ones reefers are using are off from =
the=20
factory that exact amount we are correcting for :lol:</FONT></DIV>
<DIV> </DIV>
<DIV><FONT face=3DArial size=3D2>So it appears that with the these =
Chinese refracts,=20
zeroed to RO/DI water, that they are right on when tested to a seawater =
std,=20
when they should not be. So it reads 35 ppt, 3.5 % or 1,0265 no =
adjustment is=20
needed to our surprise. Just use as is :D</FONT></DIV>
<DIV> </DIV>
<DIV><FONT face=3DArial size=3D2>Note: This is only for these refracts=20
!!</FONT></DIV>
<DIV> </DIV>
<DIV>RHS-10ATC</DIV></FONT></DIV><FONT face=3DArial size=3D2>
<DIV><A=20
href=3D"http://www.instrument-china.com/images/32-t.gif">http://www.instr=
ument-china.com/images/32-t.gif</A><BR>--=20
<BR>Boomer </DIV>
<DIV> </DIV>
<DIV>Want to talk chemistry ? The Reef Chemistry Forum<BR><A=20
href=3D"http://www.reefcentral.com/vbulletin/index.php">http://www.reefce=
ntral.com/vbulletin/index.php</A></DIV>
<DIV> </DIV>
<DIV><A=20
<BR>Former=20
US Army Bomb Technician (EOD)<BR>Member; IABTI, NATEODA, =
WEODF, ISEE=20
& IPS</DIV>
<DIV> </DIV>
<DIV>If You See Me Running You Better Catch-Up</DIV>
<DIV> </DIV>
<DIV></FONT> </DIV></BODY></HTML>
------=
|
|
Posted by Boomer on October 3, 2005, 10:22 am
Please log in for more thread options
: 7bit
I made an error on this post. The ones tested are the blue ones that most people
have,
made by Xiamen Ming Xin Instrument Company. Both
Xiamen Ming Xin Instrument Company and Huake Instrument Co., Ltd ( the black
ones) use
the same model number RHS-10ATC.
[img]http://www.mingxin-instrument.com/images/DSC04095--3.jpg [/img]
--
Boomer
Want to talk chemistry ? The Reef Chemistry Forum
http://www.reefcentral.com/vbulletin/index.php
WCWing@nospamChartermi.Net
Former US Army Bomb Technician (EOD)
Member; IABTI, NATEODA, WEODF, ISEE & IPS
If You See Me Running You Better Catch-Up
------=_NextPart_000_000B_01C5C7FB.F72D0A50
Content-Type: text/html;
charset="iso-8859-1"
Content-Transfer-Encoding: quoted-printable
<!DOCTYPE HTML PUBLIC "-//W3C//DTD HTML 4.0 Transitional//EN">
<HTML><HEAD>
<META http-equiv=3DContent-Type content=3D"text/html; =
charset=3Diso-8859-1">
<META content=3D"MSHTML 6.00.2900.2096" name=3DGENERATOR>
<STYLE></STYLE>
</HEAD>
<BODY bgColor=3D#ffffff>
<DIV><FONT face=3DArial size=3D2>I made an error on this post. The ones =
tested are=20
the blue ones that most people have, made by Xiamen Ming Xin Instrument =
Company. Both </FONT></DIV>
<DIV><FONT face=3DArial size=3D2></FONT><FONT face=3DArial =
size=3D2> Xiamen Ming=20
Xin Instrument Company and Huake Instrument Co., Ltd ( the black ones) =
use the=20
same model number RHS-10ATC.</FONT></DIV>
<DIV><FONT face=3DArial size=3D2></FONT> </DIV>
<DIV><FONT face=3DArial=20
size=3D2>[img]http://www.mingxin-instrument.com/images/DSC04095--3.jpg [/i=
mg]</FONT></DIV>
<DIV><BR>-- <BR>Boomer </DIV>
<DIV> </DIV>
<DIV>Want to talk chemistry ? The Reef Chemistry Forum<BR><A=20
href=3D"http://www.reefcentral.com/vbulletin/index.php">http://www.reefce=
ntral.com/vbulletin/index.php</A></DIV>
<DIV> </DIV>
<DIV><A=20
<BR>Former=20
US Army Bomb Technician (EOD)<BR>Member; IABTI, NATEODA, =
WEODF, ISEE=20
& IPS</DIV>
<DIV> </DIV>
<DIV>If You See Me Running You Better Catch-Up</DIV>
<DIV><FONT face=3DArial size=3D2></FONT> </DIV></BODY></HTML>
------=
|
|
Posted by Boomer on October 3, 2005, 10:24 am
Please log in for more thread options : 7bit
http://www.mingxin-instrument.com/images/DSC04095--3.jpg
--
Boomer
Want to talk chemistry ? The Reef Chemistry Forum
http://www.reefcentral.com/vbulletin/index.php
WCWing@nospamChartermi.Net
Former US Army Bomb Technician (EOD)
Member; IABTI, NATEODA, WEODF, ISEE & IPS
If You See Me Running You Better Catch-Up
I made an error on this post. The ones tested are the blue ones that most
people have,
made by Xiamen Ming Xin Instrument Company. Both
Xiamen Ming Xin Instrument Company and Huake Instrument Co., Ltd ( the black
ones) use
the same model number RHS-10ATC.
[img]http://www.mingxin-instrument.com/images/DSC04095--3.jpg [/img]
--
Boomer
Want to talk chemistry ? The Reef Chemistry Forum
http://www.reefcentral.com/vbulletin/index.php
WCWing@nospamChartermi.Net
Former US Army Bomb Technician (EOD)
Member; IABTI, NATEODA, WEODF, ISEE & IPS
If You See Me Running You Better Catch-Up
------=_NextPart_000_0027_01C5C7FC.4FE1C7D0
Content-Type: text/html;
charset="iso-8859-1"
Content-Transfer-Encoding: quoted-printable
<!DOCTYPE HTML PUBLIC "-//W3C//DTD HTML 4.0 Transitional//EN">
<HTML><HEAD>
<META http-equiv=3DContent-Type content=3D"text/html; =
charset=3Diso-8859-1">
<META content=3D"MSHTML 6.00.2900.2096" name=3DGENERATOR>
<STYLE></STYLE>
</HEAD>
<BODY bgColor=3D#ffffff>
<DIV><FONT face=3DArial size=3D2><A=20
href=3D"http://www.mingxin-instrument.com/products-refractometer.htm"></A=
></FONT> </DIV>
<DIV><FONT face=3DArial size=3D2></FONT><FONT face=3DArial size=3D2><A=20
href=3D"http://www.mingxin-instrument.com/images/DSC04095--3.jpg">http://=
www.mingxin-instrument.com/images/DSC04095--3.jpg</A></FONT><BR>--=20
<BR>Boomer </DIV>
<DIV> </DIV>
<DIV>Want to talk chemistry ? The Reef Chemistry Forum<BR><A=20
href=3D"http://www.reefcentral.com/vbulletin/index.php">http://www.reefce=
ntral.com/vbulletin/index.php</A></DIV>
<DIV> </DIV>
<DIV><A=20
<BR>Former=20
US Army Bomb Technician (EOD)<BR>Member; IABTI, NATEODA, =
WEODF, ISEE=20
& IPS</DIV>
<DIV> </DIV>
<DIV>If You See Me Running You Better Catch-Up</DIV>
<DIV> </DIV>
<DIV> </DIV>
<BLOCKQUOTE dir=3Dltr=20
style=3D"PADDING-RIGHT: 0px; PADDING-LEFT: 5px; MARGIN-LEFT: 5px; =
BORDER-LEFT: #000000 2px solid; MARGIN-RIGHT: 0px">
<DIV>"Boomer" <<A=20
=
>=20
wrote in message <A=20
=
..</DIV>
<DIV><FONT face=3DArial size=3D2>I made an error on this post. The =
ones tested are=20
the blue ones that most people have, made by Xiamen Ming Xin =
Instrument=20
Company. Both </FONT></DIV>
<DIV><FONT face=3DArial size=3D2></FONT><FONT face=3DArial =
size=3D2> Xiamen Ming=20
Xin Instrument Company and Huake Instrument Co., Ltd ( the black ones) =
use the=20
same model number RHS-10ATC.</FONT></DIV>
<DIV><FONT face=3DArial size=3D2></FONT> </DIV>
<DIV><FONT face=3DArial=20
=
size=3D2>[img]http://www.mingxin-instrument.com/images/DSC04095--3.jpg [/i=
mg]</FONT></DIV>
<DIV><BR>-- <BR>Boomer </DIV>
<DIV> </DIV>
<DIV>Want to talk chemistry ? The Reef Chemistry Forum<BR><A=20
=
href=3D"http://www.reefcentral.com/vbulletin/index.php">http://www.reefce=
ntral.com/vbulletin/index.php</A></DIV>
<DIV> </DIV>
<DIV><A=20
=
<BR>Former=20
US Army Bomb Technician (EOD)<BR>Member; IABTI, NATEODA, =
WEODF, =20
ISEE & IPS</DIV>
<DIV> </DIV>
<DIV>If You See Me Running You Better Catch-Up</DIV>
<DIV><FONT face=3DArial =
size=3D2></FONT> </DIV></BLOCKQUOTE></BODY></HTML>
------=
|
|
Posted by kryppy on October 3, 2005, 8:41 pm
Please log in for more thread options
On Sun, 2 Oct 2005 09:16:15 -0500, "Boomer"
>Recently we have been worried about reefers using and assuming that the
refracts that ever
>one has been using in this hobby should be off - 1.5 ppt or -.15 % or - .0015
SG. We have
>been telling people in short to raise the seamix "15 units" higher to = 35 ppt
( to read
>36.5), 3.5 ( to read 3.65 and 1.0265 A( to read 1,0280).
>
>These tests have been completed on a number BUT the SAME brand refracts. These
are the
>ones on the below image. Low and behold, they should read these values on a
good lab
>refract and would need these adjustments but about 95 % of the ones reefers are
using are
>off from the factory that exact amount we are correcting for :lol:
>
>So it appears that with the these Chinese refracts, zeroed to RO/DI water, that
they are
>right on when tested to a seawater std, when they should not be. So it reads 35
ppt, 3.5 %
>or 1,0265 no adjustment is needed to our surprise. Just use as is :D
>
>Note: This is only for these refracts !!
>
>RHS-10ATC
>http://www.instrument-china.com/images/32-t.gif
WooHoo, finally something I have works right.
|
|
|